Objective: To evaluate the image acquisition capability and clinical application feasibility of the Cobot-assisted automated scanning ultrasound robot in thyroid examination. Methods: This study retrospectively enrolled patients who underwent thyroid examinations at Zhongshan Hospital (Xiamen), Fudan University, between August 2024 and December 2024. Each patient sequentially received manual ultrasound scanning and automated robotic ultrasound scanning. The following parameters were compared between the two methods: thyroid dimensions measurements, longest diameter measurements of the largest thyroid nodule, examination duration, image quality scores, diagnostic concordance, and patient feedback via questionnaires regarding the robotic examination. Statistical analyses were performed using SPSS software (version 31.0). The paired t-test was used for normally distributed continuous data, and the Wilcoxon signed-rank test was used for non-normally distributed data. The intraclass correlation coefficient and Kappa coefficient were employed to evaluate consistency. A two-sided P value<0.05 was considered statistically significant. Results: All 36 patients successfully completed both examinations without adverse events. The manual and robotic scanning groups showed high consistency in thyroid dimensions measurements and the longest diameter measurements of the largest nodule (intraclass correlation coefficient values: 0.74-0.99). However, the robotic group required significantly longer examination time[(222.6±58.9) s vs (123.3±66.8) s, P<0.05]compared to the manual group. The image quality score was slightly lower in the robotic group (4.6±0.7 vs 4.9±0.4, P=0.001), though still relatively high. Diagnostic evaluation showed high concordance in thyroid nodule diagnosis and classification between the manual and robotic groups (Kappa values: 0.90-1.00). Questionnaire results indicated high patient acceptance of the robotic examination. Conclusion: The automated ultrasound scanning robot demonstrates favorable safety profiles and a high degree of patient acceptance. For thyroid ultrasound examinations, although the robotic system requires longer scanning compared to manual scanning, and its image quality scores are slightly lower, all image ratings remain within a high range. More importantly, the system shows strong consistency with manual scanning in measurements of thyroid gland dimensions, lesion characteristics, and diagnostic performance. These findings indicate significant clinical feasibility and promising application prospects for the automated robot in reducing operator workload and standardizing thyroid screening practices.